Siemens syngo.plaza SW version VA20 with Mammography license. image processing, radiological system
Recall
- Recall Number
- Z-1975-2013
- Event Number
- 65719
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 4, 2013
- Posted
- August 15, 2013
- Terminated
- March 9, 2015
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
Siemens syngo.plaza SW version VA20 with Mammography license. image processing, radiological system
Siemens issued a Customer Safety Notice to inform customers of a potential problem when using syngo.plaza. In some instances, it can happen that the laterality information in mammography images is not displayed and these images may not be properly aligned to mammography segments.
The firm, Siemens, notified their customers on March 4, 2013 and sent a "Customer Safety Advisory Notice" letter dated April 03, 2013. The letter described the product, problem and actions to be taken. The customers were instructed to always check whether the laterality information is displayed in the image text of mammography images; if not, the DICOM header is accessible and can be checked for the tag "Laterality" (0020, 0060). If a system is not able to provide the DICOM tag "Image Laterality" (0020, 0062), configuring the image text by adding the tag was proposed. If you have sold the device/equipment and it is no longer in your possession, forward this safety notice to the new owner of the device/equipment and inform us about the new owner of the device/equipment. This issue will be resolved in the syngo.plaza software version VA20D_HF04 which is planned to be released in 02/2013. Customers were also instructed to contact the local Siemens service office to plan the update. If you have any questions, contact the SYNGO CEO or Vice President Quality at +49 (9131) 84-4874 or email to: [email protected].
Nationwide distribution: FL, MA, KS, NY, OH, PA, TX and WI.