FDA Recall Terminated

Rapid Response H.U.T. (Head Up Tilt) Table. Model 1017. Powered electrophysiology tilt table. Table can tilt -15 degrees to 90 degrees (Trendelenburg). Medical Positioning Incorporated.800-593-3246. Used in clinical and hospital settings by trained radiologists.

Recall: Z-1973-2011 · Initiated February 1, 2010

Recall

Recall Number
Z-1973-2011
Event Number
58035
Firm
Medical Positioning Inc
FEI Number
1000118382
Product Code
IXR
Status
Terminated
Root Cause
Other
Initiated
February 1, 2010
Posted
April 13, 2011
Terminated
June 15, 2011
Address
1717 Washington St, Kansas City, MO, 64108-1121

Description

Rapid Response H.U.T. (Head Up Tilt) Table. Model 1017. Powered electrophysiology tilt table. Table can tilt -15 degrees to 90 degrees (Trendelenburg). Medical Positioning Incorporated.800-593-3246. Used in clinical and hospital settings by trained radiologists.

Reason

The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.

Action

An undated Urgent - Voluntary Medical Device Product Correction letter was issued on 2/1/2010 by certified mail with return receipt requested. The letter identified the model and serial numbers of the affected product. The consignees were advised to stop using the table immediately and remove it from service until further notice. Customers were to contact Mr. Joe Hurtig at 1-800-593-3246 or [email protected] to confirm receipt of the notice and to provide the firm with the correct contact information. Individualized correction packages including replacement parts and instructions were sent to each consignee. The instructions consisted of (1) Rapid Response HUT Motor Replacement Instructions; (2) Control Box Replacement Instructions; and (3) ILD External Power Supply Removal Instructions.

Distribution

Worldwide Distribution -- USA, including Puerto Rico, and Canada.

Quantity

58 total