Alaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bonded Texium Closed Male Luer With Priming Cap, REF: 10321213T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium Closed Male Luer With Priming Cap, REF: 22000-B007T; Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, REF: 22600-0007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22603-B007T; Bonded Texium Closed Male Luer with priming cap Back Check Valve 2 SmartSite Y-Sites, REF: 24010-0007T; SmartSite Bag Access Non-Vented 0.2 MicronFilter Bonded Texium Closed Low Sorbing Tubing (PE Lined) Male Luer with priming cap SmartSite Y-site, REF: 24301-0007T; SmartSite Bag Access Non-Vented Low Sorbing Texium Back Check Valve 2 SmartSite Y-Sites, REF: 24601-B007T. BD 31-IN SEC SET 20DP W/BAG ACCESS PORT TEXIUMTM AND HANGER LOW SORBING, REF: 70001B-07T; BD 20-IN EXTENSION SET W/0.2 MICRON FILTER ONE SmartSite VALVE AND Texium, REF: 20350ET; BD Secondary Set 20 Drop Vented Non Vented With Texium & Hanger, REF: 40000-07T; BD 20 DROP SECONDARY SET SmartSite VALVE BAG ACCESS W/TEXIUM CLOSED MALE LUER HANGER, REF: 10013364T; BD 32-IN GRAV SET MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, REF: 4030B-07T; BD Texium Smartsite MICROBORE EXTENSION SET 60 IN, REF: 22005E-07T
Recall
- Recall Number
- Z-1970-2025
- Event Number
- 96737
- FEI Number
- 3012798826
- Product Code
- FPA
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- May 12, 2025
- Posted
- June 12, 2025
- Address
- BD SWITZERLAND SARL Terre Bonne Park A4 Route De Crassier 17 Eysins Switzerland
Description
Alaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bonded Texium Closed Male Luer With Priming Cap, REF: 10321213T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium Closed Male Luer With Priming Cap, REF: 22000-B007T; Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, REF: 22600-0007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22603-B007T; Bonded Texium Closed Male Luer with priming cap Back Check Valve 2 SmartSite Y-Sites, REF: 24010-0007T; SmartSite Bag Access Non-Vented 0.2 MicronFilter Bonded Texium Closed Low Sorbing Tubing (PE Lined) Male Luer with priming cap SmartSite Y-site, REF: 24301-0007T; SmartSite Bag Access Non-Vented Low Sorbing Texium Back Check Valve 2 SmartSite Y-Sites, REF: 24601-B007T. BD 31-IN SEC SET 20DP W/BAG ACCESS PORT TEXIUMTM AND HANGER LOW SORBING, REF: 70001B-07T; BD 20-IN EXTENSION SET W/0.2 MICRON FILTER ONE SmartSite VALVE AND Texium, REF: 20350ET; BD Secondary Set 20 Drop Vented Non Vented With Texium & Hanger, REF: 40000-07T; BD 20 DROP SECONDARY SET SmartSite VALVE BAG ACCESS W/TEXIUM CLOSED MALE LUER HANGER, REF: 10013364T; BD 32-IN GRAV SET MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, REF: 4030B-07T; BD Texium Smartsite MICROBORE EXTENSION SET 60 IN, REF: 22005E-07T
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.
On 5/12/2025, removal notices were mailed or emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Nurse Managers, and Distributors who were asked to do the following: 1. Immediately review your inventory for affected product. Destroy all unused affected products subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the Customer Response Form via email to [email protected] 4. In addition, distributors were asked to identify all customers that purchased any affected product and to provide a copy of the customer letter to all customers and to advise them of the removal notification. Report any complaints experienced with the use of this product to the firm's North American Regional Complaint Center: Phone: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT, Email: [email protected]
Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.
2,024,780