FDA Recall Open, Classified

Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.

Recall: Z-1969-2026 · Initiated April 14, 2026

Recall

Recall Number
Z-1969-2026
Event Number
98779
Firm
Philips North America
FEI Number
3006648320
Product Code
LNH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 14, 2026
Posted
April 27, 2026
Address
222 Jacobs St, Cambridge, MA, 02141-2296

Description

Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.

Reason

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Action

Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.

Distribution

Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.

Quantity

15 units