FDA Recall
Open, Classified
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-69 (Fix)
Recall: Z-1968-2025
·
Initiated August 10, 2023
Recall
- Recall Number
- Z-1968-2025
- Event Number
- 96874
- Firm
- Turncare, Inc
- FEI Number
- 3013682457
- Product Code
- FNM
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- August 10, 2023
- Posted
- June 12, 2025
- Address
- 230 W Parkway, Ste 6, Pompton Plains, NJ, 07444-1060
Description
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-69 (Fix)
Reason
A software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
Action
Between November 2023 and January 2024 software updates to resolve the issue were performed by TurnCare on affected units. Successful update of the software was verified on each affected controller.
Distribution
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut.
Quantity
1142 units