Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
Recall
- Recall Number
- Z-1967-2026
- Event Number
- 98779
- Firm
- Philips North America
- FEI Number
- 3006648320
- Product Code
- LNH
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- April 14, 2026
- Posted
- April 27, 2026
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2296
Description
Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
14 units