Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-59 (Fix)
Recall
- Recall Number
- Z-1967-2025
- Event Number
- 96873
- Firm
- Turncare, Inc
- FEI Number
- 3013682457
- Product Code
- FNM
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- November 1, 2023
- Posted
- June 12, 2025
- Address
- 230 W Parkway, Ste 6, Pompton Plains, NJ, 07444-1060
Description
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-59 (Fix)
A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
Between June 2023 and November 2023 software updates to resolve the issue were performed by TurnCare on affected units. The software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy. Successful update of the software was verified on each affected controller.
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut
1081 units