FDA Recall Terminated

Dragonfly Laryngeal Surface Electrode, ET Tube Size (mm): 7.5-10.0, Channel 1, Reference Numbers LSE4246MCA075 and LSE500M. for continuous EMG monitoring of the larynx during surgical procedures.

Recall: Z-1967-2021 · Initiated April 8, 2021

Recall

Recall Number
Z-1967-2021
Event Number
87999
Firm
Spes Medica Via Buccari
FEI Number
3011534533
Product Code
ETN
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
April 8, 2021
Terminated
May 10, 2022
Address
21 Genoa Italy

Description

Dragonfly Laryngeal Surface Electrode, ET Tube Size (mm): 7.5-10.0, Channel 1, Reference Numbers LSE4246MCA075 and LSE500M. for continuous EMG monitoring of the larynx during surgical procedures.

Reason

The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

Action

An email titled "Object: Spes Medica Safety Notice for possible Non-Conformities to Steril Milano" dated 4/8/21 was sent to customers. Following the communications of Steril Milano and the Ministry of Health (http:ljwww.salute.gov.it/portale/news/p3 2 1 1 1.jsp ?lingua=italiano&menu=notizie&p=dalministero&id=5365) on possible deviations of the parameters/processes defined for ethylene oxide sterilization, already validated and routinely used by Steril Milano; the company Spes Medica S.r.I. informs you that in the years 2019-2020-2021 has used the Steril Milano supplier only once in- August 2020. However only for safety reason If you have the following products in your stock, we ask you to immediately block the sale and the use and, in case, to forward this request to your customers or sub-distributors. Please contact our Customer Service to send back to us the products, we will replace with another LOT number. For any request for clarification or support, do not hesitate to contact: Eng. Giorgio Facco e-mail: [email protected] Telephone number: +39 010 80 667 67

Distribution

Distribution in the following US states: CA, MA, and SC.

Quantity

20 boxes