FDA Recall Terminated

GE Healthcare Precision 500D and Advantx Legacy/Legacy-D Radiographic and Fluoroscopic Systems. A universal diagnostic imaging system for radiographic and fluoroscopic examinations. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Legacy and Legacy D Table is intended for use in a diagnostic X-ray system to support patient during general purpose radiological procedures in the horizontal, vertical and Trendelenburg positions.

Recall: Z-1966-2015 · Initiated June 12, 2015

Recall

Recall Number
Z-1966-2015
Event Number
71572
Firm
GE Healthcare
FEI Number
2126677
Product Code
JAA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
June 12, 2015
Posted
July 2, 2015
Terminated
December 21, 2015
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Precision 500D and Advantx Legacy/Legacy-D Radiographic and Fluoroscopic Systems. A universal diagnostic imaging system for radiographic and fluoroscopic examinations. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Legacy and Legacy D Table is intended for use in a diagnostic X-ray system to support patient during general purpose radiological procedures in the horizontal, vertical and Trendelenburg positions.

Reason

The SFD/IDD safety mechanism may not engage properly at the lower range of SFD/IDD travel. This mechanism is installed to hold and prevent the SFD/IDD from falling due to a counterweight cable failure. The lower range of travel is 1.1 in (28mm) from the lowest point of normal SFD/IDD vertical compression. A fall of the SFD/IDD could result in an injury to a patient or operator.

Action

GE Healthcare sent an "Urgent Medical Device Correction" letter GEHC Ref#10902 dated June 12, 2015 to Hospital Administrators/Risk Managers, Managers of Radiology/Cardiology and Radiologists/Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution - US Nationwide including PR, GU and DC and the countries of: Afghanistan, Argentina, Australia, Austria, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Czech Republic, Dominican Republic, Ecuador, Egypt, France, Germany, Ghana, Guatemala, Honduras, Indonesia, Ireland, Israel, Italy, Jamaica, Jordan, Korea (Republic of), Kuwait, Libyan Arab Jamahiriya, Malaysia, Malta, Mexico, New Zealand, Panama, Romania, Russia, Saudi Arabia, Spain, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Vietnam, and Yemen.

Quantity

4,053 (3,804 US; 249 OUS)