FDA Recall Terminated

Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D

Recall: Z-1965-2019 · Initiated July 2, 2019

Recall

Recall Number
Z-1965-2019
Event Number
83271
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
IYN
Status
Terminated
Root Cause
Device Design
Initiated
July 2, 2019
Posted
July 12, 2019
Terminated
October 5, 2020
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D

Reason

Potential for a transducer mis-alignment in certain transvaginal probes.

Action

GE Healthcare notified customers on about 07/02/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that they could continue to use the affected probes, but prior to performing a biopsy, customers should make sure they follow the instructions for safe use in the Basic User Manual to ensure proper device alignment, specifically including the biopsy, biopsy safety and biopsy setup sections. GE Healthcare will correct all affected devices at no cost and a GE Healthcare representative will contact customers to arrange for the correction. Customers were also instructed to complete and return the Medical Device Notification Acknowledgement Response form. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Distribution in the US to Texas. International distribution to France, Greece and Switzerland.

Quantity

2 devices