2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600
Recall
- Recall Number
- Z-1962-2025
- Event Number
- 96825
- Firm
- Baxter Healthcare Corporation
- FEI Number
- 1417572
- Product Code
- ERA
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- May 9, 2025
- Posted
- June 11, 2025
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600
The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/09/2025 via USPS First Class mail. The notice explained the issue, hazard involved, and requested to cease all use of the device and return it to Baxter. Please contact Baxter Technical Support to arrange for return and replacement of your affected battery or Welch Allyn PocketScope handle at 800-535-6663 between the hours of 8:00 am and 8:00 pm Eastern Time, Monday through Friday. Press option 2, then select option 1 or 2 for language preference, then press option 1 for physical exam, and then press 3 for physical exam products. Forward a copy of the notice to any other departments with the institution that may use the affected product. Distributors were directed to notify their consignees.
US Nationwide distribution.
277 units