Terumo Advanced Perfusion System 1 Electronic O2 Blender/Analyzer. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas.
Recall
- Recall Number
- Z-1962-2012
- Event Number
- 61231
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DTQ
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- June 22, 2012
- Posted
- July 9, 2012
- Terminated
- February 19, 2013
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586
Description
Terumo Advanced Perfusion System 1 Electronic O2 Blender/Analyzer. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas.
Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.
Terumo CVS sent an "URGENT MEDICAL DEVICE RECALL-SAFETY ADVISORY" letter dated June 22, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions regarding this recall.
Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
1095 Units