FDA Recall Terminated

Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held.

Recall: Z-1961-2012 · Initiated June 19, 2012

Recall

Recall Number
Z-1961-2012
Event Number
62346
Firm
Stryker Endoscopy
FEI Number
2936485
Product Code
GEI
Status
Terminated
Root Cause
Process control
Initiated
June 19, 2012
Posted
July 9, 2012
Terminated
October 23, 2012
Address
5900 Optical Ct, San Jose, CA, 95138-1400

Description

Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held.

Reason

The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.

Action

Stryker Endoscopy sent a Urgent Device Correction letter dated June 19, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to: 1. Discard any old Instructions for Use for your Arthroscope and Hardware Set Tray. 2. Download the new IFU from the URL link: http://stryker.com/arthrotray . Complete and sign the enclosed acknowledgement of receipt form and fax to (408) 754-8378 or scan in and email to [email protected].* It is important to send the self addressed confirmation form back to Stryker. Stryker is committed to delivering the highest quality products to its customers. We sincerely apologize for any inconvenience this action may cause. If you have any questions regarding this letter please contact us at 1-800-624-4422 or via email at [email protected]. .

Distribution

Worldwide Distribution--USA (nationwide)

Quantity

6388 units