FDA Recall Terminated

Dental chairs distributed under several brand names such as DCI Equipment; Marus MaxStar; Marus NuStar; and Pelton & Crane. Part number for the bolt: 022R065 Product is labeled in part: DCI Equipment, Mfg By: Dental Equipment, LLC. DC1235, Marus Dental, Mfg By: Dental Equipment, LLC. DC1490, DC1690, and DC1700, and DC1702, Pelton & Crane, Spirit SP17.

Recall: Z-1958-2011 · Initiated March 1, 2011

Recall

Recall Number
Z-1958-2011
Event Number
58122
Firm
Dental Equipment LLC
FEI Number
1000523896
Product Code
NRU
Status
Terminated
Root Cause
Process control
Initiated
March 1, 2011
Posted
April 12, 2011
Terminated
September 21, 2011
Address
705 S Springbrook Rd, Bldg B200, Newberg, OR, 97132-7057

Description

Dental chairs distributed under several brand names such as DCI Equipment; Marus MaxStar; Marus NuStar; and Pelton & Crane. Part number for the bolt: 022R065 Product is labeled in part: DCI Equipment, Mfg By: Dental Equipment, LLC. DC1235, Marus Dental, Mfg By: Dental Equipment, LLC. DC1490, DC1690, and DC1700, and DC1702, Pelton & Crane, Spirit SP17.

Reason

Link arm assembly that mounts the rear mounted accessories to the dental chair may potentially break and fall off the dental chair.

Action

On 03/08/11, Dental Equipment LLC. started sending out the Medical Device Recall notification letter via USPS 1st class mail to US consignees and Federal Express mail to foreign consignees. The firm also sent a recall notification letter to consignees who only purchased dental accessories such as Rear Mounted Cuspidor with Coverset, Arm Articulating Rear Mounted Cuspidor - Marus Gray, Maxstar Orbit Cuspidor utility W/ Porcelain bowl, and zinc plated bolts. The letter identified the affected dental chair as Marus brand with the following model numbers: DC1490, DC1690, DC1700 and DC1702. The firm's corrective action including the replacement of the affected link arm assembly with a new link arm assembly. The firm's field representative will contact each consignee for an appointment to replace the affected link arm assembly. Consignees can contact the Dental Equipment LLC at 503-537-3617 for any questions.

Distribution

Worldwide distribution to USA, Australia, Canada, Ireland, South Africa, and Vietnam.

Quantity

55 chairs and 11 replacement bolts distributed to US and foreign consignees outside the US.