FDA Recall Terminated

Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.

Recall: Z-1955-2009 · Initiated July 17, 2009

Recall

Recall Number
Z-1955-2009
Event Number
52718
Firm
Draeger Medical Systems, Inc.
FEI Number
2510954
Product Code
FMT
Status
Terminated
Root Cause
Equipment maintenance
Initiated
July 17, 2009
Posted
August 25, 2009
Terminated
June 1, 2011
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.

Reason

Hot particles fell into the infant bassinet beneath a Radiant Infant Warmer. The resulting fire injured an infant. These devices are obsolete.

Action

On 7/20/09, Draeger Medical, the recalling firm, issued a press release to medical media outreach organizations and an Urgent Medical Device letter to all known customers. This information explained the incident and the need to remove and replace the units in question as soon as feasibly possible.

Distribution

Worldwide distribution: USA, Brazil, Canada, China, Dominican Republic, France, Korea, Lebanon, Mexico, Nigeria, Panama, Poland, Saudi Arabia, Taiwan, Turkey, United Kingdom, Uruguay, and Venezuela.

Quantity

6489 units