FDA Recall Terminated

Treatment Planning and Delivery System Software version 3.6. ViewRay. Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-1954-2015 · Initiated May 7, 2015

Recall

Recall Number
Z-1954-2015
Event Number
71505
Firm
Viewray Incorporated
FEI Number
3007546534
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
May 7, 2015
Posted
July 2, 2015
Terminated
February 4, 2016
Address
2 Thermo Fisher Way, Oakwood Village, OH, 44146-6536

Description

Treatment Planning and Delivery System Software version 3.6. ViewRay. Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Reason

The software was failing to determine new patient locations if imaging is not enabled during treatment.

Action

On May 7, 2015 the firm sent Customer Advisory Notification Letters to their customers.

Distribution

Distributed in the states of CA, MO & WI.

Quantity

3