FDA Recall Terminated

ISE Electrolyte Reference reagent (for Synchron LX Systems and UniCel DxC 600/800 Systems), Part No. A28937, A28945, 467935, 467915. Product Usage: ISE Electrolyte Buffer reagent and ISE Electrolyte Reference reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1, 2 and 3, are intended for the quantitative determination of potassium concentration in human serum, plasma or urine.

Recall: Z-1951-2013 · Initiated September 22, 2010

Recall

Recall Number
Z-1951-2013
Event Number
65655
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
CEM
Status
Terminated
Root Cause
Process control
Initiated
September 22, 2010
Posted
August 10, 2013
Terminated
August 10, 2013
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

ISE Electrolyte Reference reagent (for Synchron LX Systems and UniCel DxC 600/800 Systems), Part No. A28937, A28945, 467935, 467915. Product Usage: ISE Electrolyte Buffer reagent and ISE Electrolyte Reference reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1, 2 and 3, are intended for the quantitative determination of potassium concentration in human serum, plasma or urine.

Reason

Beckman Coulter is recalling the Synchron Systems LX20/UniCel DxC ISE Reference Reagent because it has a potassium (K) response near the instrument calibration flag limits and may exceed the limits on some instruments.

Action

Beckman Coulter Inc sent an Urgent Product Corrective Action (PCA) letter dated September 22, 2010 to all affected customers. Customers were notified of the the problems identified and the actions to be completed. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact Customer Support Center (Call Center) on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support/ or call (800) 854-3633 in the US and in Canada.

Distribution

Worldwide Distribution - USA Nationwide and the countries of Belgium, France, Germany, Netherlands, Poland, Spain, Sweden, Switzerland, and United Kingdom

Quantity

718 units total (132 units in US)