FDA Recall Open, Classified

HA FlexTrak II-Patient transport functionality to transport the patient from the preparation area to the MR system, Product Number: 989710008732

Recall: Z-1950-2024 · Initiated May 3, 2024

Recall

Recall Number
Z-1950-2024
Event Number
94603
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
LNH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 3, 2024
Posted
June 4, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

HA FlexTrak II-Patient transport functionality to transport the patient from the preparation area to the MR system, Product Number: 989710008732

Reason

Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard

Action

Philips NA issued URGENT Medical Device Correction letter to US consigneess dated May 3, 2024 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take: A. If you see hydraulic fluid (oil) leakage: " First remove fluid from the floor using a paper towel. Then remove any remaining oil film from the floor using a cleaning agent (e.g. soap and water) and paper towel. " Contact your local Philips service representative. B. You may continue to use your HA FlexTrak Trolley in accordance with the intended use. C. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm until this issue has been resolved. D. Please post this letter near the associated MR system(s) for ease of reference. E. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction, understanding of the issue, and required actions to be taken. A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the HA FlexTrack Trolley and correct if required (reference FCO78100580). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday).

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lithuania, Martinique, Mexico, Morocco, Netherlands, Norway, Oman, Peru, Poland, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.

Quantity

590 units