FDA Recall Terminated

Quickie/Zippie powered wheelchair. Q700M wheelchair. A battery operated powered wheelchair, which is controlled by Power Module RNET 120A Expandable (Item No.: 250ER10, Part No.: 123966).

Recall: Z-1947-2020 · Initiated February 13, 2020

Recall

Recall Number
Z-1947-2020
Event Number
85431
Firm
Sunrise Medical (US) LLC
FEI Number
2937137
Product Code
ITI
Status
Terminated
Root Cause
Software Design Change
Initiated
February 13, 2020
Posted
May 11, 2020
Terminated
April 28, 2021
Address
2842 N Business Park Ave, Fresno, CA, 93727-1328

Description

Quickie/Zippie powered wheelchair. Q700M wheelchair. A battery operated powered wheelchair, which is controlled by Power Module RNET 120A Expandable (Item No.: 250ER10, Part No.: 123966).

Reason

Due to programming errors in the Controller, an increase in the set motor parameters can be made by the end user beyond tested safe limits.

Action

An URGENT: MEDICAL DEVICE RECALL letter was sent the firm's dealer. The dealer was instructed on February 4, 2020, to replace the controller unit for the device.

Distribution

Distributed to US Consignee in MT.

Quantity

1