IsoGel AIR 84, Support Surface. Intended to assist in improving human patient outcomes.
Recall
- Recall Number
- Z-1946-2013
- Event Number
- 65337
- Firm
- Stryker Medical Division of Stryker Corporation
- FEI Number
- 1831750
- Product Code
- IOQ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 12, 2013
- Posted
- August 9, 2013
- Terminated
- September 11, 2014
- Address
- 3800 E Centre Ave, Portage, MI, 49002-5826
Description
IsoGel AIR 84, Support Surface. Intended to assist in improving human patient outcomes.
Complaints were received from the field indicating that some Air Pumps were flashing and alarming when connected to the IsoGel AIR" surface upon turning the pump on and the setting it to the IsoGel AIR Mode. Upon further investigation, it was determined that the flashing/alarm being experienced was due to the air pressure being out of the expected range; indicating a blockage within the LAL tubi
On 7/12/13, an URGENT MEDICAL DEVICE RECALL NOTIFICATION was mailed to consignees. The notification addressed the affected product and issue, potential hazard, actions needed including faxing (269-488-8691) or emailing ([email protected]) the attached reply form to confirm receipt of notification, and risk mitigation. Please contact the firm at 269-389-8129 if you have questions or concerns regarding this recall. Business hours are M-F 8am-5pm(ET).
Nationwide Distribution-USA including the states of OH, ME, IL, CO , FL, TX, MI, GA, and VA.
586 units