FDA Recall Terminated

IsoGel AIR 84, Support Surface. Intended to assist in improving human patient outcomes.

Recall: Z-1946-2013 · Initiated July 12, 2013

Recall

Recall Number
Z-1946-2013
Event Number
65337
Firm
Stryker Medical Division of Stryker Corporation
FEI Number
1831750
Product Code
IOQ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 12, 2013
Posted
August 9, 2013
Terminated
September 11, 2014
Address
3800 E Centre Ave, Portage, MI, 49002-5826

Description

IsoGel AIR 84, Support Surface. Intended to assist in improving human patient outcomes.

Reason

Complaints were received from the field indicating that some Air Pumps were flashing and alarming when connected to the IsoGel AIR" surface upon turning the pump on and the setting it to the IsoGel AIR Mode. Upon further investigation, it was determined that the flashing/alarm being experienced was due to the air pressure being out of the expected range; indicating a blockage within the LAL tubi

Action

On 7/12/13, an URGENT MEDICAL DEVICE RECALL NOTIFICATION was mailed to consignees. The notification addressed the affected product and issue, potential hazard, actions needed including faxing (269-488-8691) or emailing ([email protected]) the attached reply form to confirm receipt of notification, and risk mitigation. Please contact the firm at 269-389-8129 if you have questions or concerns regarding this recall. Business hours are M-F 8am-5pm(ET).

Distribution

Nationwide Distribution-USA including the states of OH, ME, IL, CO , FL, TX, MI, GA, and VA.

Quantity

586 units