FDA Recall Terminated

Safedraw Venous Close Loop System, Blood Sampling Set, packaged in a tyvek bag, Model/Part Number 688695, Argon Critical Care Systems, Singapore Pte. Ltd., 198 Yishun Avenue 7, Singapore 768028. The blood sampling set is intended to be used for closed-loop blood sampling from the patient without the system being exposed to the outside environment.

Recall: Z-1946-2012 · Initiated May 4, 2012

Recall

Recall Number
Z-1946-2012
Event Number
61832
Firm
Argon Medical Devices, Inc
FEI Number
1625425
Product Code
DSK
Status
Terminated
Root Cause
Packaging
Initiated
May 4, 2012
Posted
July 6, 2012
Terminated
January 14, 2013
Address
1445 Flat Creek Rd, Athens, TX, 75751-5002

Description

Safedraw Venous Close Loop System, Blood Sampling Set, packaged in a tyvek bag, Model/Part Number 688695, Argon Critical Care Systems, Singapore Pte. Ltd., 198 Yishun Avenue 7, Singapore 768028. The blood sampling set is intended to be used for closed-loop blood sampling from the patient without the system being exposed to the outside environment.

Reason

Some of the kits may exhibit a potential defect with the pouch seal. The pouch issue may compromise the sterile barrier.

Action

Argon Medical Devices sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 4, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Recall Response Form was attached to the letter for customers to complete and return to the firm via fax to 903-577-9393. Contact the firm at [email protected] for questions regarding this recall.

Distribution

Distributed only in Missouri.

Quantity

10 pieces