FDA Recall Terminated

syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573.

Recall: Z-1944-2014 · Initiated May 20, 2014

Recall

Recall Number
Z-1944-2014
Event Number
68394
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JQP
Status
Terminated
Root Cause
Device Design
Initiated
May 20, 2014
Posted
June 30, 2014
Terminated
August 10, 2015
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573.

Reason

Certain configuration parameters may cause the reference ranges to not hold results as expected and release them to the Laboratory Information System (LIS) that should have been held for manual review.

Action

An urgent medical device correction notice, dated May 20, 2014, was sent to end users to provide instructions to avoid potential risks associated with the software defect, until changes can be made by a Siemens technical representative.

Distribution

Worldwide Distribution. US Nationwide, Canada, Australia, Belgium, Czech Republic, Germany, Denmark, Finland, France, United Kingdom, Italy, Netherlands, Portugal, Slovakia, Singapore, Japan, and Spain.

Quantity

336