Terumo Cardiovascular Systems Corporations, CDI 100 Hematocrit/ Oxygen Saturation Monitoring System. In-line blood-gas sensor used in cardiopulmonary bypass.
Recall
- Recall Number
- Z-1944-2011
- Event Number
- 57189
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DTY
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 1, 2010
- Posted
- April 11, 2011
- Terminated
- April 17, 2012
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586
Description
Terumo Cardiovascular Systems Corporations, CDI 100 Hematocrit/ Oxygen Saturation Monitoring System. In-line blood-gas sensor used in cardiopulmonary bypass.
Inaccurate result. Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.
On 11/1/2010, Terumo sent "URGENT MEDICAL DEVICE CORRECTION" notices to Chief Perfusionist, Department of Cardiovascular Surgery or Director of Operating Room Services. The letter informed the consignees about the incidents of the CDI 100 and 101 displaying inaccurate values. The letter states that when a displayed value is significantly different from expectations based on the clinical situation, the user should verify the accuracy by independent means before treating. The letter included "Caution" stickers that were to be applied to the CDI system operators manual.
US, India, Colombia, Taiwan, Venezuela, Vietnam, Puerto Rico, Malaysia, Sri Lanka, Brazil, Indonesia, Pakistan, Chile, Germany, Hong Kong, Japan, AE, Singapore, Austria, Belgium and Thailand.
1032 units