FDA Recall Terminated

Terumo Cardiovascular Systems Corporations, CDI 100 Hematocrit/ Oxygen Saturation Monitoring System. In-line blood-gas sensor used in cardiopulmonary bypass.

Recall: Z-1944-2011 · Initiated November 1, 2010

Recall

Recall Number
Z-1944-2011
Event Number
57189
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DTY
Status
Terminated
Root Cause
Device Design
Initiated
November 1, 2010
Posted
April 11, 2011
Terminated
April 17, 2012
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Terumo Cardiovascular Systems Corporations, CDI 100 Hematocrit/ Oxygen Saturation Monitoring System. In-line blood-gas sensor used in cardiopulmonary bypass.

Reason

Inaccurate result. Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.

Action

On 11/1/2010, Terumo sent "URGENT MEDICAL DEVICE CORRECTION" notices to Chief Perfusionist, Department of Cardiovascular Surgery or Director of Operating Room Services. The letter informed the consignees about the incidents of the CDI 100 and 101 displaying inaccurate values. The letter states that when a displayed value is significantly different from expectations based on the clinical situation, the user should verify the accuracy by independent means before treating. The letter included "Caution" stickers that were to be applied to the CDI system operators manual.

Distribution

US, India, Colombia, Taiwan, Venezuela, Vietnam, Puerto Rico, Malaysia, Sri Lanka, Brazil, Indonesia, Pakistan, Chile, Germany, Hong Kong, Japan, AE, Singapore, Austria, Belgium and Thailand.

Quantity

1032 units