FDA Recall Open, Classified

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Recall: Z-1942-2026 · Initiated March 25, 2026

Recall

Recall Number
Z-1942-2026
Event Number
98554
Firm
Ventec Life Systems, Inc.
FEI Number
3013095415
Product Code
NOU
Status
Open, Classified
Root Cause
Process change control
Initiated
March 25, 2026
Posted
May 1, 2026
Address
22002 26th Ave Se, Bothell, WA, 98021-4903

Description

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Reason

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

Action

On 3/25/2026 recall notices were mailed to customers who were asked to do the following: 1. Immediately discontinue use of affected devices. 2. Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3. Use a non-affected ventilator for patient support. 4. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5. Complete and return the response form via email to [email protected] 6. If you are no longer the owner of any of these devices, you need to complete an ownership transfer form at: VentecLife.com/tracking so firm can follow up with the current owner of the impacted device(s). If you have questions, you can contact the firm at 844-698-6276

Distribution

US: OH, GA, CA, MO

Quantity

42