Inova Diagnostics BINDAZYME Human C1q binding Circulating Immune Complex EIA Kit, Part#: MK021
Recall
- Recall Number
- Z-1942-2010
- Event Number
- 55801
- Firm
- Inova Diagnostics Incorporated
- FEI Number
- 2026994
- Product Code
- DAK
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 28, 2009
- Posted
- June 30, 2010
- Terminated
- June 30, 2010
- Address
- 9900 Old Grove Rd, San Diego, CA, 92131-1638
Description
Inova Diagnostics BINDAZYME Human C1q binding Circulating Immune Complex EIA Kit, Part#: MK021
The recall was initiated after the binding site notified Inova Diagnostics, inc. that the affected Bindazyme Human C1q binding Circulating Immune Complex EIA Kits may contain an ELISA plate labeled as "tissue Transglutaminase Coated Wells" instead of "C1q Coated Wells".
Notification letters from INOVA Diagnostics Incorporated were mailed to each of the consignees via Certified mail on October 30, 2009. The notification letters inform the consignees of the reason for recall and provided instructions for final disposition of the recalled kits. In addition, a FAX back Attachment was attached to the recall notice for customers to respond to the recall.
Worldwide distribution: USA, Germany, Italy, Belgium, Romania, Portugal, Argentina, Spain, Sweden, and India.
700 kits