FDA Recall Terminated

Victory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.

Recall: Z-1941-2012 · Initiated November 26, 2010

Recall

Recall Number
Z-1941-2012
Event Number
57350
Firm
Pride Mobility Products Corp
FEI Number
3005068474
Product Code
INI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 26, 2010
Posted
July 3, 2012
Terminated
March 22, 2013
Address
182 Susquehanna Ave, Exeter, PA, 18643

Description

Victory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.

Reason

The main harness connector was manufactured without the required back shield sealing protection. As a result, there is the potential for water to seep through the nonconforming harness.

Action

Three (3) pieces of information was distributed in the notification. They are: (1) Letter of Correction to the Provider (2) Letter of Correction for the Consumer (provider to issue to consumer) (3) Product Insert for Victory 9 Controller Interface Harness Replacement. The letter of Correction for Victory 9 with Power Sear (SC609PS), dated November 2010, identified the affected product and the reason for the correction. Customers were to record the serial number(s) and contact their provider's office to make an appointment to have the harness replaced.

Distribution

Worldwide Distribution

Quantity

186