FDA Recall
Terminated
STERIS 4085, 5085 & 5085SRT Surgical Tables
Recall: Z-1941-2010
·
Initiated May 14, 2010
Recall
- Recall Number
- Z-1941-2010
- Event Number
- 55850
- Firm
- Steris Corporation
- FEI Number
- 1527821
- Product Code
- FQO
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 14, 2010
- Posted
- June 30, 2010
- Terminated
- February 24, 2012
- Address
- 5960 Heisley Rd, Mentor, OH, 44060
Description
STERIS 4085, 5085 & 5085SRT Surgical Tables
Reason
STERIS has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. This supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table
Action
On May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.
Distribution
Worldwide distribution: USA, Australia, India, Korea & New Zealand
Quantity
79 units