FDA Recall Terminated

Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.

Recall: Z-1940-2020 · Initiated April 15, 2020

Recall

Recall Number
Z-1940-2020
Event Number
85456
Firm
Cook Inc.
FEI Number
1820334
Product Code
NIN
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 15, 2020
Terminated
April 15, 2024
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.

Reason

The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.

Action

On April 15, 2020, the firm distributed Urgent Medical Device Recall letters to affected consignees via email. The letter described the product defect and advised that customers do the following: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving the letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. The recall notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. Cook Medical is removing impacted devices from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue.

Distribution

Domestic distribution to OH, IA, VA, and NC. Foreign distribution to Canada.

Quantity

10