BD Biosciences, BD FACS Sample Prep Assistant II (SPA II), blood sample preparer for flow cytometry analysis
Recall: Z-1940-2010
·
Initiated February 19, 2010
Recall
- Recall Number
- Z-1940-2010
- Event Number
- 55654
- Firm
- BD Biosciences, Systems & Reagents
- FEI Number
- 2916837
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- February 19, 2010
- Posted
- June 30, 2010
- Terminated
- May 18, 2011
- Address
- 2350 Qume Dr, San Jose, CA, 95131-1812
Description
BD Biosciences, BD FACS Sample Prep Assistant II (SPA II), blood sample preparer for flow cytometry analysis
Reason
Samples prepared on the BD SPA II, SN T0076 produced inaccurate sample results due to inaccurate sample dispensed as a result of the cap piercing probe
Action
The firm filed an MDR and implemented a field action consisting of a February 2010 notification letter and product replacement of affected probes. A CAPA for the probe design has been implemented. Effectivity is being tracked via mail and follow up phone call.
Distribution
Worldwide distribution: USA, Australia, Belgium, China (Hong Kong), Colombia, Jamaica, Korea, Mexico, Singapore and Uruguay.
Quantity
349 units