FDA Recall Terminated

SOMATOM go.Now, Material Number 11061618 There is a potential for a software issue that may cause the need for necessary patient rescans.

Recall: Z-1939-2018 · Initiated April 20, 2018

Recall

Recall Number
Z-1939-2018
Event Number
80047
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
April 20, 2018
Terminated
May 23, 2019
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

SOMATOM go.Now, Material Number 11061618 There is a potential for a software issue that may cause the need for necessary patient rescans.

Reason

There is a potential for a software issue that may cause the need for necessary patient rescans.

Action

The firm, Siemens Healthineers, sent an "Customer Safety Advisory Notices" letter dated 4/20/2018 beginning on 4/20/18 to its customers. The letter described the product, problem and actions to be taken. The letters provided the following information to customers: How can the operator help to avoid a potential risk of the system? To avoid an unnecessary additional X-ray dose to patients please observe the following instructions: Topic 1: If a reconstruction cannot be started after an examination, please reboot the system and repeat the aborted reconstruction. If the error persists, please contact your local Siemens Healthineers service support team. They can help you to decide whether a rescan is necessary or not. Topic 2: If you change the scan parameters kV or mA (tube HV or current), please wait for at least 12 seconds before you trigger a scan command on the control box. If artifacts are visible in the topogram, please repeat the topogram before starting the volume scan. If you have started a volume scan and detect artifacts in the real time display, please stop the scan immediately to avoid an unnecessary X-ray dose. Topic 3: After an osteoporosis scan has been completed, the Osteo reconstruction needs to be started and completed directly after the scan. Therefore the images should be loaded and checked in CT View&Go before the scan/recon tab card is closed. Please do not close the scan/recon tab card before you have reconstructed and checked all Osteo images. A successful and complete Osteo reconstruction is visualized by full green bars in the column Recon Progress of the Job Status table and by images including the Osteo phantom in the tab card CT View&Go (for details please see Fig. 3 below). An example of a correct Osteo reconstruction is provided in Fig. 2. After checking the successful and complete Osteo reconstruction you may close the scan/recon tab card. What has been done to address these issues? What actions shall be t

Distribution

Worldwide Distribution: US (nationwide) including states of: AR, CA, CT, FL, GA, IL, IN, KS, KY, MO, ND, NY, OH, OK, SD, TN, TX, and VA; and countries of: Algeria, Australia, Austria, Bangladesh, Bolivia, Bosnia Herzog., Brazil, Cambodia, Chile, China, Columbia, Costa Rica, Germany, Greece, India, Indonesia, Italy, Japan, Kenya, Kuwait, Lithuania, Mauritius, Morocco, Nepal, Pakistan, Paraguay, Philippines, Poland, Portugal, Republic Korea, Romania, Senegal, Singapore, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, Ukraine, and United Kingdom.

Quantity

4 units in US