Maestro Air Motor Pneumatic cranial drill motor. The Maestro Air Pneumatic System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories.

Recall: Z-1939-2016 · Initiated April 28, 2016

Recall

Recall Number
Z-1939-2016
Event Number
74170
Firm
Stryker Instruments Div. of Stryker Corporation
FEI Number
1811755
Product Code
HBB
Status
Terminated
Root Cause
Component design/selection
Initiated
April 28, 2016
Posted
June 9, 2016
Terminated
March 24, 2017
Address
4100 E Milham Ave, Portage, MI, 49002-9704

Description

Maestro Air Motor Pneumatic cranial drill motor. The Maestro Air Pneumatic System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories.

Reason

Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. The grease may also migrate into the attachment end of the motor. There is a negligible possibility that the orange grease may leak from the attachment end of the motor and contaminate the surgical wound resulting in soft tissue inflammation.

Action

Stryker sent an "Urgent Medical Device Recall Notification" dated April 28, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, customer responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments at (269) 389-4518.

Distribution

Domestic only: AL AR CA FL GA IL IN KS KY MD MI MO NC NM OH OK SC TN TX VA WA WI

Quantity

189