FDA Recall Terminated

EXPRESS Dry Seal Chest Drain, Single Collection P/N: 4000-100N

Recall: Z-1938-2015 · Initiated May 7, 2015

Recall

Recall Number
Z-1938-2015
Event Number
71023
Firm
Atrium Medical Corporation
FEI Number
3011175548
Product Code
PAD
Status
Terminated
Root Cause
Process control
Initiated
May 7, 2015
Posted
June 29, 2015
Terminated
April 26, 2016
Address
5 Wentworth Dr, Hudson, NH, 03051-4929

Description

EXPRESS Dry Seal Chest Drain, Single Collection P/N: 4000-100N

Reason

Chest drains have a missing o-ring

Action

Atrium Medical issued recall letter dated 5/7/15 to inform customers of the steps to take should they encounter affected product. If the O-ring is not present on the patient tubing: a. Do not use the chest drain. b. Replace the entire chest drain including the patient tubing. Open up another chest drain package and repeat steps 1 to 3. c. Return any units with missing O-rings to Atrium. If the O-ring is present then reconnect in-line connector and proceed with set up as normal. Complete and return Field Action Reply Form to the following e-mail address: [email protected] or you may fax it to 1-603-386-6590. If you have any questions, please contact your local Atrium/Maquet chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST).

Distribution

Nationwide Canada, Germany

Quantity

390 cases