BD BBL CultureSwab EZ Collection and Transport System, REF 220144, in vitro Diagnostic, SterileR, Rx Only
Recall
- Recall Number
- Z-1937-2025
- Event Number
- 96807
- Firm
- Copan Italia Via Francesco Perotti 10
- FEI Number
- 3002444944
- Product Code
- JTW
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 2, 2025
- Posted
- June 9, 2025
- Address
- 18 Brescia Italy
Description
BD BBL CultureSwab EZ Collection and Transport System, REF 220144, in vitro Diagnostic, SterileR, Rx Only
Swabs for specimen collection may be prone to breakage during product usage.
On May 2, 2025, Copan Italia Spa issued a Urgent: Medical Device Recall" notification to affected consignees via E-Mail. Copan asked consignees to take the following steps: 1. Trace and dispose of all affected products 2. Disposal must be carried out according to local regulations. Evidence or declaration of Product disposal must be returned in writing 3. This notice should be passed on to all those who need to be aware within your organization. if you have further distributed the devices, please provide those Customers a copy of the recall notification and response form. 4. Confirm you have read and understood the recall instructions provided in the letter and submit the acknowledgment form.
US Nationwide distribution in the state of MD.
81,600 units