FDA Recall Terminated

iChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland, Made in Germany.

Recall: Z-1937-2019 · Initiated October 18, 2018

Recall

Recall Number
Z-1937-2019
Event Number
82879
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
KQO
Status
Terminated
Root Cause
Component change control
Initiated
October 18, 2018
Terminated
November 10, 2020
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

iChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland, Made in Germany.

Reason

Falsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction.

Action

The recalling firm issued letters dated 10/15/2018 on 10/18/2018 via email or regular mail for customers where an email was unavailable. The letter explains the issue and impact, and provides the actions the consignee is to take when using the product. The letter informed the customer they are actively working on a resolution to prevent recurrence of the issue.

Distribution

Distribution was nationwide, including Puerto Rico. There was also military/government distribution. Foreign distribution was made to Canada, Argentina, Brazil, Chile, Colombia, Mexico, Peru, Philippines, Singapore, and Taiwan.

Quantity

956,798/100 test strip vials