FDA Recall Terminated

EASYMax N. Self-Monitoring Blood Glucose System, distributed in cardboard carton kits. Each kit contains the following components: Blood Glucose Meter; Test Strip x 10 ct.; Normal Control Solution; Carrying Case; Alkaline Battery x 2 ct.; Lancets x 5 ct.; Lancing Device; AST Lancing Device Cap; User's Manual; Test Strip Instructions and Control Solution Instructions. Indicated for use at home by persons with diabetes or in clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control.

Recall: Z-1937-2009 · Initiated May 26, 2009

Recall

Recall Number
Z-1937-2009
Event Number
52352
Firm
Oak Tree International Holdings, LLC, Inc.
FEI Number
3007644653
Product Code
NBW
Status
Terminated
Root Cause
Process control
Initiated
May 26, 2009
Posted
September 10, 2009
Terminated
September 27, 2012
Address
1209 Lowell St, Ste 10, Elyria, OH, 44035-4803

Description

EASYMax N. Self-Monitoring Blood Glucose System, distributed in cardboard carton kits. Each kit contains the following components: Blood Glucose Meter; Test Strip x 10 ct.; Normal Control Solution; Carrying Case; Alkaline Battery x 2 ct.; Lancets x 5 ct.; Lancing Device; AST Lancing Device Cap; User's Manual; Test Strip Instructions and Control Solution Instructions. Indicated for use at home by persons with diabetes or in clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control.

Reason

Blood glucose meters, which were set to the mmol/L. unit of measure, rather than the accepted/labeled mg./dL unit of measure, were distributed.

Action

EPS Technology issued a certified notification letter to the sole distributor/customer: GEMCO, Coconut Creek, FL dated May 26, 2009. The letter describes the potential preset error and its possible consequences to the user/customer. The letter goes on to inform the distributor that EPS Technology Corp. will be sending out letters to all affected user/customers during a direct mailing. If the error is found, the user/customer can either: 1) have the firm's Customer Service Dept. instruct them as to how to fix the unit, or 2) request that a replacement meter be sent to them. For further information, contact EPS Technology Customer Service at 1-866-994-3345.

Distribution

Nationwide Distribution -- AL, AR, AZ, CA, CT, FL, GA, IL, IN, KS, KY, LA, MD, ME, MI, MS, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, VA, WA, WY and HI.

Quantity

5,396 units