FDA Recall Open, Classified

Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., polyethylene or nylon, non-braided or braided with 1:1 torque). The choice of catheter material for different angiographic procedures should be based on the physician s experience.

Recall: Z-1934-2026 · Initiated April 2, 2026

Recall

Recall Number
Z-1934-2026
Event Number
98682
Firm
Cook Incorporated
FEI Number
1820334
Product Code
DQO
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 2, 2026
Posted
May 1, 2026
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., polyethylene or nylon, non-braided or braided with 1:1 torque). The choice of catheter material for different angiographic procedures should be based on the physician s experience.

Reason

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

Action

On April 2, 2026, the firm began notifying affected consignees via letters titled "URGENT: MEDICAL DEVICE RECALL." Customers were instructed to quarantine affected products in inventory and to immediately cease further distribution and use of the affected products. They should return the affected products to Cook Medical with a copy of the completed Acknowledgement and Receipt form to receive a product credit.

Distribution

US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay

Quantity

11,253