FDA Recall Terminated

SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Recall: Z-1934-2015 · Initiated June 4, 2015

Recall

Recall Number
Z-1934-2015
Event Number
71484
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
June 4, 2015
Posted
June 27, 2015
Terminated
March 24, 2017
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Reason

A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.

Action

A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.

Distribution

Nationwide Distribution.

Quantity

40 total