FDA Recall Open, Classified

RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator.

Recall: Z-1933-2024 · Initiated April 16, 2024

Recall

Recall Number
Z-1933-2024
Event Number
94533
Firm
Reflexion Medical, Inc.
FEI Number
3011716550
Product Code
QVA
Status
Open, Classified
Root Cause
Software design
Initiated
April 16, 2024
Posted
May 30, 2024
Address
25881 Industrial Blvd, Hayward, CA, 94545-2991

Description

RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator.

Reason

A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.

Action

On 04/29/2024, the firm sent an "Urgent: Medical Device Recall Safety Notification" via email to customer informing them that during internal dosimetric testing of SCINTIX OOS partials in periodic tracking mode, it was observed that the delivered dose error of the full fraction can exceed 5% with respect to the upper bound of one or more static volumes within the DVH in the plan (non-tracked, non-target volumes in the patient frame of reference, i.e., the planning CT). This was determined to be the result of a software defect associated with completing interrupted. Customers are instructed to: 1. Notify all clinical staff of this recall 2. Utilized the Clinical Guidance Document that is include with this notification. For questions or concerns - contact Service and Support at 650-239-9070 or email [email protected]

Distribution

U.S.: CA, CT, NJ, OR, PA and TX No O.U.S.

Quantity

7 systems