FDA Recall
Terminated
SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
Recall: Z-1933-2015
·
Initiated June 4, 2015
Recall
- Recall Number
- Z-1933-2015
- Event Number
- 71484
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 4, 2015
- Posted
- June 27, 2015
- Terminated
- March 24, 2017
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
Reason
A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
Action
A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.
Distribution
Nationwide Distribution.
Quantity
40 total