Medtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the system voltage): Product Name Model#/ CFN OARM ASSY 8170000027 SYSTEM PRODUCT BI70000027 OARM ASSY BI70000027R SYS PRODUCT RWK BI70000027R BASE OARM 8170000027100 SYSTEM !DOV BI70000027100 BASE OARM Bl70000027100R SYSTEM lOOV RWK Bl70000027100R BASE OARM 8170000027120 SYSTEM 120V 8170000027120 BASE OARM Bl70000027120R SYSTEM 120V RWK BI70000027120R BASE OARM Bl70000027230 SYSTEM 230V 8170000027230 BASE OARM Bl70000027230R SYSTEM 230V RWK BI70000027230R OARM ASSY BI70000027GER SYS PRODUCT GER BI70000027GER OARM ASSY Bl70000027GERR SYS PRODUCT REFURB BI70000027GERR BASE OARM 8170000028100 SYS lOOV 8170000028100 BASE OARM BI70000028120 SYS 120V 8170000028120 BASE OARM BI70000028120R 3RD EDIT REFURB BI70000028120R BASE OARM BI70000028230 SYS 230V BI70000028230 BASE OARM BI70000028230R 3RD EDIT REFURB Bl70000028230R Product Usage: The 0-arrn Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomi~ structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Recall
- Recall Number
- Z-1932-2019
- Event Number
- 82991
- Firm
- Medtronic Navigation, Inc.-Littleton
- FEI Number
- 3004785967
- Product Code
- OXO
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 1, 2019
- Terminated
- January 19, 2021
- Address
- 300 Foster St, Littleton, MA, 01460-2017
Description
Medtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the system voltage): Product Name Model#/ CFN OARM ASSY 8170000027 SYSTEM PRODUCT BI70000027 OARM ASSY BI70000027R SYS PRODUCT RWK BI70000027R BASE OARM 8170000027100 SYSTEM !DOV BI70000027100 BASE OARM Bl70000027100R SYSTEM lOOV RWK Bl70000027100R BASE OARM 8170000027120 SYSTEM 120V 8170000027120 BASE OARM Bl70000027120R SYSTEM 120V RWK BI70000027120R BASE OARM Bl70000027230 SYSTEM 230V 8170000027230 BASE OARM Bl70000027230R SYSTEM 230V RWK BI70000027230R OARM ASSY BI70000027GER SYS PRODUCT GER BI70000027GER OARM ASSY Bl70000027GERR SYS PRODUCT REFURB BI70000027GERR BASE OARM 8170000028100 SYS lOOV 8170000028100 BASE OARM BI70000028120 SYS 120V 8170000028120 BASE OARM BI70000028120R 3RD EDIT REFURB BI70000028120R BASE OARM BI70000028230 SYS 230V BI70000028230 BASE OARM BI70000028230R 3RD EDIT REFURB Bl70000028230R Product Usage: The 0-arrn Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomi~ structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
Medtronic Navigation issued letter May l, 2019 via FedEx Express Overnight delivery identifying the issue, health risk and action to take: provided a visual mitigation card to attach to the Image Acquisition System (IAS) of their affected systems to serve as a visual reminder. " Consignees will be asked to sign and return the consignee response form; acknowledging receipt of this information and confirmation of having attached the visual mitigation card to their affected system. Review and retain Attachment 8 for further guidance . Attachment B provides instructions for inspecting the battery and chargers. and follow the "Performance Checks and Maintenance" section of your user manual, in addition to the information in Attachment B, to ensure your system is available for surgical procedures Medtronic field service personnel will install a new version of the user manual " Medtronic field service personnel will install a component (Mobile View Station \'MVS'') heatsink) as part of the new design of the AC power input circuit. Questions contact Medtronic Technical Support if you have any questions or concerns regarding this matter: USA: (800) 595-9709 or 720-890-3160 Worldwide: +1-720-890-3160 RS.NA VTECHSUPPORT@M EDTRON IC.COM
Worldwide Distribution - US Nationwide Foreign: ALGERIA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL BULGARIA CANADA CANADA CHILE CHINA COLOMBIA CROATIA CZECH REPUBLIC DENMARK EGYPT FINLAND FRANCE GERMANY GREECE HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KOREA KOREA, REPUBLIC OF KUWAIT LEBANON LIBYAN ARAB JAMAHIRIYA LK MEXICO NETHERLANDS NEW ZEALAND NICARAGUA POLAND PORTUGAL PUERTO RICO RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SLOVAKIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM VENEZUELA
832 units