FDA Recall Terminated

Uric Acid, Catalog No. OSR6x98 Product Usage:Uric acid OSR6x98 is intended as a System reagent for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on Beckman Coulter AU analyzers.

Recall: Z-1932-2016 · Initiated April 13, 2016

Recall

Recall Number
Z-1932-2016
Event Number
73921
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
KNK
Status
Terminated
Root Cause
Device Design
Initiated
April 13, 2016
Terminated
June 30, 2016
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Uric Acid, Catalog No. OSR6x98 Product Usage:Uric acid OSR6x98 is intended as a System reagent for the quantitative determination of Uric Acid in human serum, heparinized plasma and urine on Beckman Coulter AU analyzers.

Reason

Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.

Action

Beckman Coulter sent an Urgent Medical Device Recall letter dated April 13, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions regarding the customer notification letter are instructed to contact Customer Support Center at http://www.beckmancoulter.com/customersupport/support; (800) 223-0130 in the US and Canada. Outside of the US and Canada, customers are instructed to contact their local Beckman Coulter Representative.

Distribution

Worldwide Distribution - US Nationwide including Puerto Rico and the countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guam, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Lebanon, Lesotho, Liberia, Libyan Arab Jamahiriya, Lithuania, Malawi, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Zambia, and Zimbabwe.

Quantity

59,401 units total (10,868 units in US)