FDA Recall Open, Classified

FREESTYLE LIBRE 2 GLUCOSE MONITORING SYSTEM: Item number: 5658877

Recall: Z-1931-2021 · Initiated March 16, 2021

Recall

Recall Number
Z-1931-2021
Event Number
87787
Firm
Cardinal Health Inc.
FEI Number
3004008697
Product Code
QLG
Status
Open, Classified
Root Cause
Environmental control
Initiated
March 16, 2021
Address
7000 Cardinal Pl, Dublin, OH, 43017-1091

Description

FREESTYLE LIBRE 2 GLUCOSE MONITORING SYSTEM: Item number: 5658877

Reason

Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.

Action

On 03/16/2021, Cardinal Health issued an Urgent Medical Device Removal notice to customer via letter notifying them a recall is being initiated as a precautionary measure as a result of exposure to temperatures outside of the products labeling requirements. Actions Required 1. INSPECT your inventory for the affected product. 2. SEGREGATE and QUARANTINE all affected product. 3. RETURN the enclosed acknowledgment form via fax to 1-817-868-5362 or email to [email protected], even if you do not have any product on hand. 4. CONTACT Inmar to arrange for return and credit of any affected product: 1-855-696-2074 (9:00 AM to 5:00 PM, Monday thru Friday)

Distribution

US Nationwide distribution in the states of FL, GA, SC.

Quantity

2,115 kits