MEDLINE Sterile* Needle Counter, 100Ct. Double Magnet / Foam Block, reorder: DYNJNC100F Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.
Recall
- Recall Number
- Z-1931-2018
- Event Number
- 80017
- Firm
- Medline Industries Inc
- FEI Number
- 1417592
- Product Code
- GAB
- Status
- Terminated
- Root Cause
- Package design/selection
- Initiated
- April 24, 2018
- Terminated
- August 4, 2020
- Address
- Three Lakes Drive, Northfield, IL, 60093
Description
MEDLINE Sterile* Needle Counter, 100Ct. Double Magnet / Foam Block, reorder: DYNJNC100F Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.
The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters.
Medline sent an Urgent Recall letter dated April 24, 2018. The letter identified the affected product, problem and actions to be taken. Customer were instructed to check inventory, quarantine all affected product and return completed enclosed response form. Those consignees who further distributed the product were directed to notify their customers. For question contact 866-359-1704.
Worldwide Distribution - US Nationwide and the countries of: Canada, Saudi Arabia, Cayman Island, and United Arab Emirates
14208 units