FDA Recall Terminated

Constellation Vision System, Model: Constellation LX, Catalogue Number: 8065751551 The Constellation Vision System is an ophthalmic microsurgical system that is indicated for both anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e., vitreoretinal) ophthalmic surgery in addition to the indications included with the optional Next Generation laser.

Recall: Z-1930-2010 · Initiated April 30, 2010

Recall

Recall Number
Z-1930-2010
Event Number
55648
Firm
Alcon Research LTD dba Alcon Laboratories, Inc.
FEI Number
1000125771
Product Code
HQC
Status
Terminated
Root Cause
Software design
Initiated
April 30, 2010
Posted
July 13, 2010
Terminated
September 19, 2011
Address
15800 Alton Pkwy, Irvine, CA, 92618-3818

Description

Constellation Vision System, Model: Constellation LX, Catalogue Number: 8065751551 The Constellation Vision System is an ophthalmic microsurgical system that is indicated for both anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e., vitreoretinal) ophthalmic surgery in addition to the indications included with the optional Next Generation laser.

Reason

Alcon identified system performance and machine settings that may impact the infusion performance of the CONSTELLATION Vision System, causing unexpected shut down, unresponsive touch screens, and fluidics issues.

Action

The firm, Alcon, sent a "URGENT: MEDICAL DEVICE RECALL" letter dated July 2, 2010, along with training material to all customers. The letter informed the customers of the problem and indicated that Alcon Surgical Sales representative will be contacting them in the near future to follow-up on this training material and to answer any questions you may have. Alcon informed the customers that this action will not require removal of the CONSTELLATION Vision System from their facility. The letter also informed the customers that their Alcon Technical Customer Service Representative will be contacting them when we are authorized to provide you with the software and hardware updates for your system. For assistance and questions regarding the CONSTELLATION Vision System, customers were told they may contact their Alcon Surgical Sales Representative or Alcon Technical Consumer Affairs at 1-800-561-6466, Option 8.

Distribution

Worldwide distribution.

Quantity

824 units total