FDA Recall Terminated

VP 2000 Processor; an automated tissue stainer 100 volt - part 30-144100 or list 2J11-60; 117 volt - part 30-144101 or list 2J11-01, and part 30-102832 - processor only; 230 volt - part 30-144102 or list 2J11-04, and part 30-102833 processor only. Designed to automate and standardize slide specimen processing including deparafinization, FISH pretreatment and routine slide staining for the laboratory. The unit is preprogrammed to perform slide pretreatment protocols for FISH assays and can be programmed to perform routine histology/cytology slide staining or routine slide washing according to user requirements.

Recall: Z-1929-2009 · Initiated September 15, 2008

Recall

Recall Number
Z-1929-2009
Event Number
52055
Firm
Abbott Molecular
FEI Number
3005248192
Product Code
KEY
Status
Terminated
Root Cause
Device Design
Initiated
September 15, 2008
Posted
August 30, 2009
Terminated
January 7, 2010
Address
1300 E Touhy Ave, Des Plaines, IL, 60018-3315

Description

VP 2000 Processor; an automated tissue stainer 100 volt - part 30-144100 or list 2J11-60; 117 volt - part 30-144101 or list 2J11-01, and part 30-102832 - processor only; 230 volt - part 30-144102 or list 2J11-04, and part 30-102833 processor only. Designed to automate and standardize slide specimen processing including deparafinization, FISH pretreatment and routine slide staining for the laboratory. The unit is preprogrammed to perform slide pretreatment protocols for FISH assays and can be programmed to perform routine histology/cytology slide staining or routine slide washing according to user requirements.

Reason

Thermal runaway conditions, in that the target temperature was exceeded and the temperature display could not track the actual temperature within the heated slide basins of the VP 2000 Slide Processor.

Action

Abbott sent Device Recall/Urgent Field Safety Notice letters dated 9/15/08 to the VP2000 customers on 9/12/08, informing them that thermal runaway conditions may occur within the heated slide basins of the VP2000 Slide Processor. The customers were given actions to take to prevent this condition: a) VP2000 should be powered down when not in use and when the controlling computer is not in use; b) Liquid level in all heated basins should be checked periodically to ensure that there is adequate reagent present for slide processing and to keep the sensor probe immersed for accurate temperature sensing; and c) If large, rapid and random fluctuations in indicated basin temperature occurs the unit should be shut down and a Field Service Representative be contacted. Any questions were directed to their local Abbott Technical Service Representative.

Distribution

Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Columbia, Czech Republic, Egypt, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Korea, Malaysia, Norway, Russia, Saudi Arabia, Singapore, Switzerland and the United Kingdom.

Quantity

261 units