FDA Recall Open, Classified

CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179

Recall: Z-1928-2018 · Initiated May 3, 2018

Recall

Recall Number
Z-1928-2018
Event Number
80003
Firm
Maquet Datascope Corp Cardiac Assist Division
FEI Number
3001418283
Product Code
DSP
Status
Open, Classified
Root Cause
Device Design
Initiated
May 3, 2018
Address
1300 Macarthur Blvd, Mahwah, NJ, 07430-2052

Description

CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179

Reason

Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.

Action

Notification of the field correction was posted to the firm's website on 5/3/18 at https://www.maquet.com/globalassets/downloads/product-articles/cardiosave/ml_0730_us_rev_a_cardiosave_fluid_ingress_recall_customer_letter_us_4may2018.pdf. Mailed notification letters will be issued to customers as well. The instructions included in the notification includes the following: Patients receiving IABP therapy are in critical condition. Failure to start or sudden interruption of therapy could result in unsafe, hemodynamic instability. Until the protective top cover is installed, please adhere to the following instructions when using Cardiosave intra-aortic balloon pump: 1) Pursuant to the Caution section of our Cardiosave IABP Operating/User Instructions, Never place fluids on top of this unit. Make sure that the saline container and tubing do not hang directly over the IABP. In case of accidental spillage, wipe clean immediately and have the unit serviced to ensure no hazard exists In the unlikely event that a sudden interruption of therapy occurs, transfer the patient to an alternative IABP. The Intra-Aortic Balloon (IAB) Catheter Instructions for Use reiterates that a catheter should not remain inactive for more than 30 minutes, due to the potential for thrombus formation. If an alternative IABP is unavailable; manually inflate the IAB with air or helium and immediately aspirate, repeat every 5 minutes until either an alternate IABP is available or alternatively, the intra-aortic balloon catheter should be removed from the patient. Please refer to the intra-aortic balloon catheter instructions for use, Manually Inflating and Deflating a Catheter. The patient should be treated according to your facilitys treatment protocols and caregivers clinical judgment to ensure hemodynamic stability. Affected customers will be contacted by a Maquet/Getinge representative to schedule on-site service of your Cardiosave IABP by a Maquet/ Getinge Sales or Service Re

Distribution

The devices were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Guam. The devices were distributed to the following foreign countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, Chile, China, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrica, Taiwan, Thailand, Turkey, UAE, UK, Vietnam, Yemen.

Quantity

4300