Medtronic Bio-Medicus Femoral Cannula/Introducer, for Cardiopulmonary Bypass. Catalog Numbers: CB96345-023, CB96345-023, and CB96345-029.
Recall
- Recall Number
- Z-1928-2008
- Event Number
- 48100
- Firm
- Medtronic Perfusion Systems
- FEI Number
- 1000116158
- Product Code
- DWF
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 9, 2008
- Posted
- July 24, 2008
- Terminated
- May 8, 2012
- Address
- 7611 Northland Dr N, Brooklyn Park, MN, 55428-1088
Description
Medtronic Bio-Medicus Femoral Cannula/Introducer, for Cardiopulmonary Bypass. Catalog Numbers: CB96345-023, CB96345-023, and CB96345-029.
Medtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it had provided some batches to Medtronic that were contaminated with oversulfated chondroitin sulfate (OSCS).
Medtronic issued a Press Release on 05/07/2008. An Urgent Medical Device Recall letter was sent to customers on May 07, 2008. The letter identifies affected devices, explains the situation and requests that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall Certificate acknowledging the receipt of this information.
Worldwide Distribution: USA, Australia, Belgium, Canada, Czech Republic, France, Italy, Japan, Norway, and Taiwan.
17 US, 29 OUS