FDA Recall Terminated

FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.

Recall: Z-1925-2012 · Initiated February 20, 2012

Recall

Recall Number
Z-1925-2012
Event Number
62118
Firm
Maquet Cardiovascular Us Sales, Llc
FEI Number
3008355164
Product Code
BSZ
Status
Terminated
Root Cause
Device Design
Initiated
February 20, 2012
Posted
July 6, 2012
Terminated
September 9, 2015
Address
45 Barbour Pond Drive, Wayne, NJ, 07470

Description

FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.

Reason

A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.

Action

The firm, Maquet Getinge Group, sent an "URGENT DEVICE FIELD CORRECTION" letter dated February 20, 2012 to its customers. The letter describes the product, problem and action to be taken. The letter state that a Maquet Service Representative will upgrade the customers software in the FLOW-i Anesthesia System to the current revision and also provide the customers with the corresponding user's manual. Note: This upgrade includes other minor improvements. The customers were instructed to sign a document to verify satisfactory completion of the work once the software upgrade was completed. If you have questions or require additional information, please contact your local MAQUET representative or Technical Support at 888-627-8383. A press release was issued on 7/6/2012.

Distribution

Worldwide distribution: USA (nationwide) state of GA only and countries of: Argentina, Australia, Austria, Belgium, China, Czech Republic, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, Spain, Sweden incl Maquet Critical Care, Switzerland, Thailand, United Arab Emirates, and United Kingdom.

Quantity

284 units