FDA Recall Terminated

Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part Number 010066. Top Assembly Part Numbers: 040906, 040904, 040756, 040758, 040658, 040680. Product Usage: Device is used to provide treatment for neonatal hyperbilirubinemia.

Recall: Z-1924-2017 · Initiated April 3, 2017

Recall

Recall Number
Z-1924-2017
Event Number
74474
Firm
Natus Medical Incorporated
FEI Number
3018859
Product Code
LBI
Status
Terminated
Root Cause
Component change control
Initiated
April 3, 2017
Posted
April 14, 2017
Terminated
August 13, 2019
Address
5900 1st Ave S, Seattle, WA, 98108-3248

Description

Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part Number 010066. Top Assembly Part Numbers: 040906, 040904, 040756, 040758, 040658, 040680. Product Usage: Device is used to provide treatment for neonatal hyperbilirubinemia.

Reason

The second generation neoBLUE 2 lights with a black band, sold from 2003 to 2008, are recalled so it will eliminate the possibility that a neoBLUE system may have a replacement board installed which allows it to be used at a higher overall light intensity than the devices were originally cleared to deliver.

Action

On April 3, 2017, Natus sent the URGENT: RECALL NOTIFICATION 2nd Generation neoBLUE LED Phototherapy Systems letter, dated March 2017, and the Reply Form to the U.S. customers. U.S. Customers are advised to follow four steps process to receive replacement lights and return affected light back to Natus in Seattle, WA for destruction. Customers should scan and email completed Reply Form to [email protected] or FAX to 847-949-6031. On April 4, 2017, Natus sent the URGENT MEDICAL DEVICE RECALL/EXCHANGE 2nd Generation neoBLUE LED Phototherapy Systems letter, dated March 2017, Reply Form, Instructions to Disable neoBLUE 2 Light, neoBLUE 2 Systems Serial Number List For Their Territory to international Distribution Partners. Downstream international customers are instructed to return the affected neoBLUE lights to their Distribution Partners. International Distribution Partners follow the Instructions to Disable neoBLUE2 lights (DOC- 019983 rev A) and destroy all affected neoBLUE lights that they recover from downstream international customers. Distribution Partners can contact Natus Technical Service at 1-888-496-2887, or e-mail the Reply Form to [email protected] or via fax to 847-949-6031.

Distribution

Worldwide Distribution - US Nationwide and in the following countries: Argentina, Australia, Austria, Belgium, Bermuda, Canada, Colombia, Croatia, Denmark, Germany, Greece, Hungary, India, Indonesia, Israel, Italy, Japan, Kenya, Kuwait, Lithuania, Morocco, Norway, Peru, Philippines, Poland, Portugal, Romania, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Virgin Islands (British).

Quantity

4,371 units (3,285 units in the US and 1,086 units outside of the US)