FDA Recall Terminated

BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile non-invasive device used to connect venous access devices such as needles, blood collection sets, ad infusion sets to blood collection tubes.

Recall: Z-1923-2013 · Initiated May 1, 2013

Recall

Recall Number
Z-1923-2013
Event Number
65322
Firm
Becton Dickinson & Company
FEI Number
2243072
Product Code
JKA
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 1, 2013
Posted
August 7, 2013
Terminated
December 4, 2015
Address
1 Becton Dr, Franklin Lakes, NJ, 07417-1815

Description

BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile non-invasive device used to connect venous access devices such as needles, blood collection sets, ad infusion sets to blood collection tubes.

Reason

BD Vacutainer Multiple Sample and Direct Draw Luer Adapter may have blood leakage and low draw volume.

Action

BD sent a Product Advisory Notice dated May 3, 2013, via UPS to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discard filled tubes since it is possible they will contain less blood volume (low draws), replace the Luer Adaptor unit being used with a fresh unit, and continue to draw blood samples using the fresh unit. Customers with questions were instructed to contact BD Technical Service at 1-800-631-0174, Option 4, or contact them through website: www.bd.com/vacutainer/contact. For questions regarding this recall call 201-847-6800.

Distribution

Nationwide Distribution

Quantity

37,906,400 units