BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile non-invasive device used to connect venous access devices such as needles, blood collection sets, ad infusion sets to blood collection tubes.
Recall
- Recall Number
- Z-1923-2013
- Event Number
- 65322
- Firm
- Becton Dickinson & Company
- FEI Number
- 2243072
- Product Code
- JKA
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- May 1, 2013
- Posted
- August 7, 2013
- Terminated
- December 4, 2015
- Address
- 1 Becton Dr, Franklin Lakes, NJ, 07417-1815
Description
BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile non-invasive device used to connect venous access devices such as needles, blood collection sets, ad infusion sets to blood collection tubes.
BD Vacutainer Multiple Sample and Direct Draw Luer Adapter may have blood leakage and low draw volume.
BD sent a Product Advisory Notice dated May 3, 2013, via UPS to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discard filled tubes since it is possible they will contain less blood volume (low draws), replace the Luer Adaptor unit being used with a fresh unit, and continue to draw blood samples using the fresh unit. Customers with questions were instructed to contact BD Technical Service at 1-800-631-0174, Option 4, or contact them through website: www.bd.com/vacutainer/contact. For questions regarding this recall call 201-847-6800.
Nationwide Distribution
37,906,400 units